Search icon
×
Positioning iconCurrent Location:Home>PRODUCTS>Petrochemical Analysis Instrument>Karl Fischer Moisture Analyzer>HM-WS1 Karl Fischer Coulometric Moisture Analyzer
HM-WS1 Karl Fischer Coulometric Moisture Analyzer

HM-WS1 Karl Fischer Coulometric Moisture Analyzer

Product Number:

HM-WS1

Product Difference:

The HM-WS1 is a budget-friendly semi-automatic Karl Fischer coulometric moisture analyzer featuring an LED five-digit display, electrolysis current auto-control up to 400 mA, and 60-second automatic determination. It provides a practical entry-level solution for trace moisture measurement across industrial liquid and solid samples.

/skin/Product Details


HM-WS1 semi-automatic Karl Fischer coulometric moisture analyzer front view

HM-WS1 LED five-digit decimal display showing moisture measurement result

HM-WS1 sealed titration cell bottle for safe Karl Fischer reagent handling

HM-WS1 automatic stirring control during coulometric titration process

HM-WS1 liquid sample injection for trace water determination

HM-WS1 solid sample injector accessory for indirect moisture measurement

HM-WS1 titration cell assembly with sealed bottle and electrodes

HM-WS1 petroleum product sample being prepared for moisture analysis

HM-WS1 chemical reagent bottles for Karl Fischer coulometric titration

HM-WS1 measurement range display from 1ppm to 100 percent moisture

HM-WS1 sensitivity threshold indicator showing 0.1 microgram water detection

HM-WS1 compact benchtop design for standard laboratory workspace

HM-WS1 AC 220V power supply connection and rear panel

HM-WS1 complete instrument setup with titration cell and accessories

Product Introduction

The HM-WS1 is a semi-automatic Karl Fischer coulometric moisture analyzer designed for budget-conscious laboratories requiring reliable trace water determination. Employing the Karl Fischer coulometric titration principle — based on the quantitative reaction of iodine, sulfur dioxide, and water — the instrument measures moisture content across a range of 0.0001% (1 ppm) to 100% with a sensitivity threshold of 0.1 μg H₂O.

The instrument features automatic stirring control and completes a typical determination in approximately 60 seconds. A sealed titration cell bottle prevents reagent exposure to the operator, while compatible solid and liquid sample injectors enable indirect measurement of diverse sample types. The LED five-digit decimal display provides clear numerical readout, and the electrolysis current is automatically controlled up to a maximum of 400 mA.

Compliant with 23 national and industry standards (GB/T and SH/T series), the HM-WS1 is suitable for trace moisture measurement in petroleum, chemical, pharmaceutical, pesticide, and electric power applications.

Applications / Testing Items

  • Petroleum products moisture determination

  • Chemical industry trace water analysis

  • Pharmaceutical moisture content testing

  • Pesticide water content measurement

  • Electric power insulation oil testing

  • Liquid sample trace water determination

  • Solid sample indirect moisture measurement

Features

  1. Karl Fischer coulometric titration (coulometric analysis) based on I₂/SO₂/water quantitative reaction

  2. Semi-automatic operation with automatic stirring control for straightforward daily use

  3. Measurement range from 0.0001% (1 ppm) to 100% moisture content

  4. Sensitivity threshold of 0.1 μg H₂O for trace-level water detection

  5. Automatic determination completing in approximately 60 seconds

  6. Sealed titration cell bottle preventing reagent-operator contact

  7. Compatible with solid and liquid sample injectors for indirect measurement

  8. LED five-digit decimal display for clear numerical readout

  9. Electrolysis current automatically controlled up to 400 mA

  10. Compliance with 23 national and industry standards (GB/T, SH/T series)

  11. Operating temperature range of 5–40°C for standard laboratory environments

  12. AC 220V ±10% power supply compatible with standard laboratory mains

Main Parameters

ParameterSpecification
Titration MethodCoulometric titration (coulometric analysis)
Measurement Range0.0001% (1 ppm) to 100%
Titration Speed0.6 mg/min (maximum)
Sensitivity Threshold0.1 μg H₂O
DisplayLED five-digit decimal
Electrolysis CurrentAuto-controlled, max 400 mA
Operating Temperature5–40°C
Power SupplyAC 220V ±10%

Technical Parameters

ParameterSpecification
ModelHM-WS1
Titration MethodCoulometric titration (coulometric analysis)
Measurement Range0.0001% (1 ppm) to 100%
Titration Speed0.6 mg/min (maximum)
Sensitivity Threshold0.1 μg H₂O
Display SystemLED five-digit decimal
Electrolysis CurrentAuto-controlled, max 400 mA
Operating Temperature5–40°C
Power SupplyAC 220V ±10%
Applicable Standards23 national and industry standards (GB/T, SH/T)