The HM-KFV1 is a Karl Fischer volumetric moisture titrator designed for precise water content determination in samples with moderate to high moisture levels. Based on the volumetric titration principle, it quantifies water through the stoichiometric reaction of iodine, sulfur dioxide, and water, using dead-stop endpoint detection for accurate determination of the titration equivalence point. The instrument is verified against JJG1154-2018, the national verification regulation for Karl Fischer volumetric moisture titrators.
Built on the Android operating system with a 7-inch color touchscreen, the HM-KFV1 provides an integrated platform for titration control, data management, and report generation. It comes standard with dual burettes (10 mL and 20 mL), enabling flexible titrant volume selection with a minimum titration increment of 1 μL for precise endpoint approach. The sealed fluid path uses PTFE tubing and fluorocarbon rubber seals to prevent solvent diffusion and chemical corrosion. Compliance with FDA 21 CFR Part 11 and GLP requirements ensures data integrity in regulated environments. Multiple result units (μg, mg, %, ppm) are supported, and a built-in thermal printer provides on-demand documentation. Wi-Fi connectivity enables cloud-based firmware updates.
Applications / Testing Items
Pharmaceutical: moisture in APIs, excipients, and finished products
Chemical: water content in solvents, reagents, and intermediates
Food: moisture in oils, fats, and processed products
Petroleum: water content in lubricants, fuels, and crude oil
Liquids: direct injection of liquid samples
Gases: moisture in compressed and process gases
Soluble solids: dissolved samples in appropriate solvents
Features
Android operating system with 7-inch color touchscreen for integrated titration control
Dual burettes (10 mL and 20 mL) standard for flexible titrant volume selection
Minimum titration volume of 1 μL for precise endpoint approach
Dead-stop titration method for accurate endpoint detection
PTFE and fluorocarbon rubber sealed fluid path for anti-diffusion and corrosion resistance
FDA 21 CFR Part 11 compliance for electronic records and data integrity
GLP-compliant operation with full audit trail capability
Multiple result units display: μg, mg, %, ppm
Large storage capacity for measurement records
Built-in thermal printer for on-demand report generation
Wi-Fi connectivity for cloud-based firmware updates
Main Parameters
| Parameter | Specification |
|---|---|
| Titration Method | Volumetric (dead-stop endpoint detection) |
| Moisture Measurement Range | 10 ppm to 100% |
| Recommended Measurement Range | 0.1% to 100% |
| Measurement Resolution | 1 ppm (0.0001%) or 1 μg H₂O |
| Repeatability (RSD) | ≤1% |
| Burettes | 10 mL and 20 mL (dual standard) |
| Minimum Titration Volume | 1 μL |
| Display | 7-inch color touchscreen |
Technical Parameters
| Parameter | Specification |
|---|---|
| Model | HM-KFV1 |
| Titration Method | Karl Fischer volumetric titration, dead-stop endpoint detection |
| Moisture Measurement Range | 10 ppm to 100% |
| Recommended Measurement Range | 0.1% to 100% |
| Measurement Resolution | 1 ppm (0.0001%) or 1 μg H₂O |
| Polarization Voltage | 30 / 50 / 100 mV (±1 mV) |
| Repeatability (RSD) | ≤1% |
| Burettes | 10 mL and 20 mL (dual standard) |
| Minimum Titration Volume | 1 μL |
| Sealed Fluid Path | PTFE tubing, fluorocarbon rubber seals |
| Result Units | μg, mg, %, ppm |
| Compliance | FDA 21 CFR Part 11, GLP |
| Applicable Standard | JJG1154-2018 |
| Sample Types | Liquids, gases, soluble solids |
| Main Unit Dimensions | 310 × 226 × 233 (400) mm |
| Main Unit Weight | 3.7 kg |