Search icon
×
Positioning iconCurrent Location:Home>PRODUCTS>Petrochemical Analysis Instrument>Flash Point Tester>HM-KS2 Open Cup Flash Point and Fire Point Tester
HM-KS2 Open Cup Flash Point and Fire Point Tester

HM-KS2 Open Cup Flash Point and Fire Point Tester

Product Number:

HM-KS2

Product Difference:

HM-KS2 is an automatic open-cup flash point and fire point tester for petroleum products and other flammable liquids. It follows GB/T 3536--2008, ASTM D92, GB/T 267-88, and ISO2592-2000, using flame-ion detection to automatically lock the fire point value.

/skin/Product Details

HM-KS2 open flash point tester front view with touchscreen

Open tester features showing heating ignition self-check and thermal printing

Railway aviation power and petroleum application panels

HM-KS2 open and HM-BS2 closed tester comparison

A-grade three-wire PT100 temperature compensation on flash tester

HM-KS2 tester close-up with sample cup and ignition arm

HM-KS2 heating rate stages for open-cup testing

Open tester Android screen showing real-time temperature curve

HM-KS2 standards GB/T 3536-2008 ASTM D92 and ISO2592-2000

HM-KS2 component diagram with sensors cup touchscreen printer and USB port

HM-KS2 technical parameters compared with HM-BS2

HM Instruments team group photo in company facility

Researchers conducting laboratory testing with analytical instruments

Product Introduction

HM-KS2 is an automatic open-cup flash point and fire point tester for laboratories working with petroleum products and other flammable liquids. It heats samples according to the GB/T 3536—2008 temperature curve and performs electric ignition as the temperature approaches the preset flash point. Flame-ion detection identifies the flash, while the instrument automatically locks the fire point value when burning is detected. After the test, forced air cooling helps cool the heater. The 7-inch high-definition Android touchscreen shows the real-time temperature curve and supports Chinese-English interface switching. A-grade PT100 sensing uses a three-wire system with temperature compensation; intelligent temperature control coordinates the specified heating stages. The operating range is room temperature~400°C, with a specified repeatability of ≤4°C. Test data can be stored without a stated limit according to available EMMC space, exported, and uploaded to a cloud platform. The thermal printer supports Chinese characters, graphics, QR codes, and unlimited lines. HM Instruments specifies HM-KS2 for railway, aviation, electric power, petroleum, and scientific research applications. See HM-KS2 open-cup flash point testing for a related product view.

Applications / Testing Items

  • Petroleum product flash point and fire point testing.

  • Flammable liquid testing under the open-cup method.

  • Railway-sector material safety evaluation.

  • Aviation-sector laboratory testing and quality control.

  • Electric power laboratories requiring documented flash point data.

  • Scientific research involving applicable petroleum products and flammable liquids.

Functional Performance / Features

  1. Open-cup testing for petroleum products and other flammable liquids.

  2. Heating follows the GB/T 3536-2008 prescribed temperature curve.

  3. Automatic electric ignition is performed near the preset flash point.

  4. Flame-ion detection identifies the flash event.

  5. The fire point value is automatically locked when the burning condition is detected.

  6. A-grade PT100 with a three-wire system provides temperature compensation.

  7. Intelligent temperature control manages the specified heating stages.

  8. Forced air cooling helps cool the heater after the test.

  9. The 7-inch Android touchscreen displays a real-time temperature curve and supports Chinese-English switching.

  10. Data storage is unrestricted within available EMMC space; data export and cloud upload are provided.

  11. The thermal printer supports Chinese characters, graphics, QR codes, and unlimited lines.

  12. Self-checking covers the detection arm, sweeping rod, ignition wire, printer, and fan.

Main Parameters

ItemHM-KS2 specification
Measurement rangeRoom temperature~400°C
Temperature detectionA-grade PT100 (three-wire system, temperature compensation)
Temperature detection accuracy0.1°C
Repeatability≤4°C
Flash detection methodFlame-ion detection
Fire point functionAutomatically locks the fire point value
Ignition methodElectric ignition
Cooling methodForced air cooling

Technical Parameters

ParameterSpecification
Display7-inch high-definition Android touchscreen, 1024*600 pixels
Barometer accuracy≤0.5kPa
Heating speedBefore expected flash point 56°C: 14°C-17°C/min; starting 23°C before the expected flash point: 5°C-6°C/min
Temperature controlIntelligent temperature control
Display contentReal-time temperature curve, Chinese-English interface switching
Data storageUnlimited, according to EMMC space
Other functionsData export, data upload to cloud platform
PrinterThermal; supports Chinese characters, graphics, QR codes, with no limit on the number of lines
Self-check functionDetection arm, sweeping rod, ignition wire, printer, fan, and other checks
Execution standardsGB/T 3536--2008 Petroleum products—Determination of flash point and fire point—Cleveland open cup method; ASTM D92; GB/T 267-88; ISO2592-2000
Power supplyAC220V±10% 50Hz±5%
Power≤600W
Ambient temperature10°C~35°C
Ambient humidity<85%
Dimensions43cm*28cm*35cm (length*width*height)
Weight11.3kg

Expert Review