HM-BS2 is an automatic closed-cup flash point tester for petroleum products and chemical solvents. It follows GB/T 261-2021 and ISO2719-2016, combining A-grade three-wire PT100 temperature compensation with K-type thermocouple flash detection and automatic result recording.

/skin/Product Details

Flash point tester front view with touchscreen and sample cup

Tester feature icons for touchscreen heating ignition checking and printing

Oil electricity chemical and aviation testing application panels

HM-KS2 and HM-BS2 open and closed tester comparison

Closed flash point tester showing compensated PT100 temperature sensing

HM-BS2 instrument close-up with sample cup and touchscreen

Open and closed tester heating rate comparison graphic

HM-BS2 touchscreen showing a live temperature curve

HM-BS2 closed-cup standards GB/T 261-2021 and ISO2719-2016

HM-BS2 component diagram with sample cup lid heating tank and printer

Closed and open flash point tester technical parameter table

HM Instruments engineering team group photograph

Laboratory researchers operating scientific instruments

Product Introduction

HM-BS2 is an automatic closed-cup flash point tester for laboratories and quality-control teams that need a repeatable way to assess flammable petroleum products and chemical solvents. It heats a sample along the GB/T 261-2021 curve, ignites automatically as the temperature approaches the expected flash point, and uses a K-type thermocouple to detect the flash. When a flash is detected, the instrument locks the result, displays it, prints it, and starts fan cooling for the heater. A 7-inch high-definition Android touchscreen presents the live temperature curve and supports Chinese-English interface switching. A-grade PT100 sensing in a three-wire system provides temperature compensation, while intelligent temperature control supports the defined heating stages. Unlimited data storage depends on available EMMC space, and data can be exported or uploaded to a cloud platform. The built-in thermal printer supports Chinese characters, graphics, QR codes, and unlimited lines. HM Instruments positions this model for petroleum, chemical, electrical, inspection, and research laboratories. For a related product view, see HM-BS2 closed-cup flash point testing.

Applications / Testing Items

  • Petroleum product flash point testing in oil and energy laboratories.

  • Closed-cup testing of flammable chemical solvents.

  • Electrical-sector material safety checks where flash point data is required.

  • Inspection and commercial testing laboratories supporting storage, transport, and use decisions.

  • Scientific research work involving controlled heating and flash detection of applicable liquids.

Functional Performance / Features

  1. Closed-cup flash point measurement for petroleum products and chemical solvents.

  2. Heating follows the GB/T 261-2021 prescribed temperature curve.

  3. Automatic electric ignition is activated as the sample approaches its expected flash point.

  4. K-type thermocouple detection identifies the flash event.

  5. A-grade PT100 with a three-wire system provides temperature compensation.

  6. Intelligent temperature control supports the defined multi-stage heating rates.

  7. The result is automatically locked, displayed, and printed when a flash is detected.

  8. The 7-inch Android touchscreen displays a real-time temperature curve and supports Chinese-English switching.

  9. Data storage is unrestricted within available EMMC space; data export and cloud upload are provided.

  10. The thermal printer supports Chinese characters, graphics, QR codes, and unlimited lines.

  11. Self-checking covers the detection arm, ignition wire, printer, fan, cover switch, and related functions.

Main Parameters

ItemHM-BS2 specification
Measurement rangeRoom temperature~400°C
Temperature detectionA-grade PT100 (three-wire system, temperature compensation)
Temperature detection accuracy0.1°C
Accuracy±0.5°C
Repeatability≤3.0°C (closed flash point≤110°C); ≤5.0°C (closed flash point>110°C)
Flash detectionK-type thermocouple
IgnitionElectric ignition

Technical Parameters

ParameterSpecification
Display7-inch high-definition Android touchscreen, 1024*600 pixels
Heating speedStep A: 5~6°C/min; Step B: 1.0~1.5°C/min; Step C: 2.5~3.5°C/min
Temperature controlIntelligent temperature control
Air-bath thickness6.4mm
Display contentReal-time temperature curve, Chinese-English interface switching
Data storageUnlimited, according to EMMC space
Other functionsData export, data upload to cloud platform
PrinterThermal; supports Chinese characters, graphics, QR codes, with no limit on the number of lines
Self-check functionDetection arm, ignition wire, printer, fan, switch cover, and other checks
Execution standardsGB/T 261--2021 Flash point determination—Pensky-Martens closed cup method; ISO2719-2016
Power supplyAC220V±10% 50Hz±5%
Power≤600W
Ambient temperature10°C~35°C
Ambient humidity<85%
Dimensions43cm*28cm*35cm (length*width*height)
Weight11.7kg