Product Overview:
ORP Online Monitor is a commonly used instrument and equipment, mainly used to accurately measure the pH/redox potential value of liquid media, and is widely used in industry, agriculture, scientific research, environmental protection and other fields.
Measurement Principle:
The measuring electrode is immersed in the solution, and the H+ ions generate a battery voltage between the electrode membrane and the solution. The battery voltage generated by the electrode is proportional to the ORP value. The controller has a built-in microcomputer circuit to obtain the ORP value by measuring the battery voltage.
ORP Standard Solution Configuration
Weigh 10.2g of quinone hydroquinone and add it to 250mL of a standard buffer solution with a pH value of 4.01. Stir well and mix it into a standard solution of 256mV.
Weigh 10.2g of quinone hydroquinone and add it to 250mL of a standard buffer solution with a pH value of 6.86. Stir well and mix it into a standard solution of 86mV.
Instrument features:
❊Dot matrix LCD screen display
❊Intelligent menu operation
❊Various automatic calibration functions
❊Various signal output modes
❊Manual and automatic temperature compensation
❊Three groups of relay control switches
❊High limit, low limit, hysteresis control
❊Display pH/ORP, temperature, status, etc. on the same interface
Technical parameters:
Product name: Online PH/ORP meter
Product model: HM-ORP
Measuring range: pH: 0-14.00 ORP: -2000+2000mV Temperature: 0-150℃ (depending on the electrode)
Measurement accuracy: pH: ±0.02pH ORP: ±1mV Temperature: ±0.3℃
Temperature compensation: automatic or manual
Resolution: pH: 0.01pH ORP: 1mV Temperature: ±0.1℃
Stability: pH: ≤0.01pH/24h; ORP: ≤1mV/24h
Relay: three groups 5A 240VAC/5A 28VDC or 120VAC
Power supply: 85~265VAC or 9~36VDC, power: ≤3W
Current output: 2-way 4-20mA as standard
Communication output: RS485 Modbus RTU as standard
Dimensions: 215*125*80mm
Opening size: -
Protection level: IP65
Installation method: wall-mounted
Instrument weight: 1.5kg
Ambient temperature: -10~60℃; Relative humidity: ≤90%; No strong magnetic field interference around except the earth's magnetic field