Autosampler capacity looks like a simple specification — how many vials fit — and is actually a statement about how long the instrument can run unsupervised. A 24-position tray and a 156-position tray differ by a factor of six in sample count, but the purchase decision turns on whether the laboratory needs to load once and walk away, or can reload between batches.

This article sets out how to size that decision, using the AS-24 and AS-156 as reference points. It is written by the engineering team at HENG MEI Intelligent Manufacturing.

Start from run time, not sample count

The useful calculation is batch time: samples divided by throughput, multiplied by cycle time. A chromatographic run of 20 minutes per sample means a 24-position tray provides eight hours of unattended operation — one working day. A 156-position tray provides over 50 hours, which is more than a weekend.

PositionsAt 10 min/sampleAt 20 min/sampleAt 40 min/sample
244 h8 h16 h
15626 h52 h104 h

That table explains who buys what. A laboratory running short methods and reloading twice a day is served by 24 positions. A laboratory running long gradient methods, or wanting to load on Friday and read results on Monday, needs the larger tray — and needs to think about vial evaporation over that period.

Liquid autosampler with 156-position sample tray and control panel for chromatography systems

What the 156-position format adds beyond capacity

The AS-156 is a tower-type autosampler with a 156-position main tray and a 15-position secondary tray, designed to serve two injection ports from one instrument. That second capability is often more valuable than the raw count.

  • Two systems, one sampler. A single autosampler can feed two chromatographs or two detectors, which matters where bench space and capital budget are constrained.
  • Priority samples. The smaller tray allows a rush sample to be inserted without disturbing a loaded sequence.
  • Standard and sample separation. Standards can be placed on a dedicated tray, reducing the risk of a sequence error mid-run.
  • Host software control. The instrument is operated from a PC application, so sequences can be built and edited without touching the instrument.

Questions that decide the purchase

QuestionIf yesImplication
Does the laboratory run unattended overnight or over a weekend?Larger trayCheck vial sealing and evaporation for long sequences
Are two chromatographs served by one bench?Dual-inlet towerConfirm the second system's trigger and communication
Do rush samples arrive during a run?Secondary traySequence editing must support insertion without aborting
Are sample volumes below 100 µL?Check vial and needle depthLow volume limits how many injections per vial are possible
Is the method sensitive to carry-over?Check wash protocolAsk for the rinse sequence and the wash solvent volume

Vial handling is where automation quietly fails

An autosampler is only as reliable as the vials it picks up. Four practical issues dominate.

  1. Septum coring. Repeated piercing with a blunt needle drops fragments into the vial. Needle type and septum quality matter more than most buyers expect.
  2. Evaporation. A vial sitting in a tray for 48 hours in a volatile solvent concentrates. Long sequences need capped vials and, where available, a tray cooler.
  3. Vial height tolerances. Cheap vials vary; a needle that reaches the bottom in one vial may not in another. Specify vials from a single supplier and check the lot.
  4. Sample settling. Particulate samples settle during a long sequence; filtration before loading is cheaper than re-running a batch.

Integration and control

The autosampler must talk to whatever it feeds. Two integration questions should be settled before purchase: whether the host software can import a sequence from the laboratory's LIMS or spreadsheet, and whether the sampler provides the start signal the chromatograph expects. A sampler that requires manual sequence entry for 156 samples has removed the labour saving it promised.

For laboratories running ion chromatography or HPLC with autosamplers already installed, the practical upgrade path is usually to add capacity rather than replace the system — which makes the tray format and the control protocol the deciding specifications, not the detector.

How HENG MEI supports autosampler projects

HENG MEI Intelligent Manufacturing supplies the AS-24 and AS-156 liquid autosamplers, with the 156-position tower offering a 15-position secondary tray and dual-inlet operation, controlled from host software. Both are quoted with the laboratory's system configuration so that the trigger and sequence requirements are confirmed before delivery.

To size an autosampler, tell our sales contact the samples per batch, the cycle time per sample, whether operation must continue overnight, and which instrument the sampler feeds. Those four facts decide capacity and the integration work around it.