Search icon
×
Positioning iconCurrent Location:Home>PRODUCTS>Material Analysis Instrument>Insulation Resistance Tester>Insulation Volume Surface Resistivity Tester
Insulation Volume Surface Resistivity Tester

Insulation Volume Surface Resistivity Tester

Product Number:

HM-QDZ1

Product Price:

$1940

Product Difference:

Integrated shielding chamber and main unit with 7-inch touchscreen, automatic resistance and resistivity testing, and DC test voltages up to 1000 V.

/skin/Product Details

Volume and surface resistivity tester with integrated chamber and touchscreen

Measurement, control, output and protection features

Insulating materials and electrical resistance measurement applications

Resistance range and manufacturer-stated error bands

Integrated DC test voltage source and protective functions

Seven-inch touchscreen, automatic controls and result storage

GB and IEC standards listed in the supplied product literature

Volume and surface resistivity tester technical specifications

Hengmei company team, laboratory facilities and instrument business overview

Product Overview

The HM Instruments HM-QDZ1 volume and surface resistivity tester measures the electrical resistance and resistivity of insulating materials. It combines a shielding chamber, a DC test voltage source and a 7-inch touchscreen in one unit for laboratory material evaluation and quality control.

The specified resistance range is 5 × 10⁴ to 1 × 10¹⁷ Ω. Volume resistance, surface resistance, volume resistivity and surface resistivity are separate test results; resistance error varies with the resistance range as shown below.

Key Features

  • Measures volume resistance, surface resistance, volume resistivity and surface resistivity.

  • Integrated picoampere-sensitive current acquisition module for weak-current measurement.

  • Automatic voltage, range and timing control, with automatic calculation of results.

  • Seven-inch touchscreen displays voltage, current and resistivity in real time.

  • Built-in DC high-voltage source with six fixed voltage steps and continuously adjustable settings.

  • Real-time result printing and separate storage for three measurement modes.

  • Integrated shielding chamber and main unit, with overcurrent and short-circuit protection.

Applications

Applications include plastics, rubber, films, resins, mica, ceramics, glass, insulating varnishes and other solid insulating materials, as well as powders and liquids. Select the electrode and sample arrangement for the material and test method.

Applications include insulation-material research and quality control, polymer and composite material evaluation, university teaching and research, and electrical-property testing.

Technical Specifications

ParameterSpecification
ModelHM-QDZ1
Resistance range5 × 10⁴–1 × 10¹⁷ Ω
Resistance error: 5 × 10⁴–1 × 10⁸ Ω±1%
Resistance error: 1 × 10⁸–1 × 10¹⁰ Ω±2%
Resistance error: 1 × 10¹⁰–1 × 10¹² Ω±5%
Resistance error: 1 × 10¹²–1 × 10¹⁷ Ω±15–20%
DC test voltages10 / 50 / 100 / 250 / 500 / 1000 V; continuously adjustable setting also described in the product literature
DC voltage accuracy±5%
Power supplyAC 220 V / 10 A / 50 Hz
Power consumptionApproximately 15 W
Operating temperature0–40 °C
Operating humidity≤70%
Display7-inch touchscreen
Dimensions408 × 322.5 × 287.5 mm
WeightApproximately 9.6 kg
OutputReal-time printing of results
StorageThree modes; 18 records per mode; records can be deleted
ProtectionOvercurrent and short-circuit protection
StructureIntegrated shielding chamber and main unit

Referenced Test Standards

Method references include GB/T 1410-2006, GB/T 31838.2-2019, GB/T 31838.3-2019, GB/T 31838.4-2019 and IEC 62631, with further methods for antistatic textiles, footwear, rubber and related materials. Select the electrode arrangement and sample conditioning required by the chosen test method.

Operation and Support

Select the test mode and voltage, set the timing and relevant sample parameters, and start the automatic measurement. The touchscreen presents the measured electrical values and calculated results. Results can be stored or printed. Follow the instrument manual for sample preparation, connections and safe operation of the high-voltage test source.

Address of this article:https://www.kjhm.net/irt/723.html