Automatic Kjeldahl Analyser: Static vs Dynamic Titration | HENG MEI Intelligent Manufacturing
Article Source: Hengmei Technology Release time:2026-11-16 17:59:19
Kjeldahl nitrogen determination has three stages: digestion, distillation and titration. Automation can cover all three or only part of them, and the difference between a semi-automatic and a fully automatic analyser is not convenience — it is what happens to operator time and to the repeatability of the titration end point.
This article explains what full automation covers in the HM-DN2, how static and dynamic titration differ, and where the practical limits sit. It is written by the engineering team at HENG MEI Intelligent Manufacturing.
What "automatic" covers
Digestion happens in a separate block digestor and is not part of the analyser. Everything after digestion can be automatic: dilution, alkali addition, steam distillation, absorption into boric acid, titration, emptying and rinsing of the reaction vessel.
A semi-automatic instrument stops before titration; the operator titrates manually and enters the volume. That single manual step is where most laboratory-to-laboratory variation enters, because end-point judgement is visual and operator-dependent. An automatic titrator with a colour sensor removes that variable and records the volume, which is what makes a result auditable.

Static versus dynamic titration
Two titration modes appear across automatic analysers, and the choice affects throughput more than accuracy.
| Mode | Sequence | Effect |
|---|---|---|
| Static (HM-DN2) | Distillation completes, then titration begins | Simpler control; predictable cycle; slightly longer per sample |
| Dynamic (HM-DN3) | Titration proceeds while distillation continues | Shorter cycle at high sample loads; requires tighter control of both processes simultaneously |
For a laboratory running 20–40 samples a day, static titration is usually sufficient: at 3–9 minutes per sample, the difference is measured in minutes, not hours. At 100 or more samples a day, the overlap in dynamic mode becomes the reason to specify it.
Specification and what each limit means
| Parameter | HM-DN2 | Practical meaning |
|---|---|---|
| Measuring range | 0.1–240.0 mg nitrogen | Covers trace and high-protein samples without changing configuration |
| Sample size | Solid ≤8.0 g, liquid ≤30.0 mL | Fixed by tube volume; larger samples require sub-sampling |
| Accuracy | Relative difference ≤1% | Agreement with a reference determination |
| Repeatability | Parallel difference ≤0.5% | Agreement between duplicate digestions — the number laboratories quote to clients |
| Titration precision | 1.0 µL per step | Resolution of the delivered titrant volume |
| Recovery | 99–101% | Distillation and absorption efficiency, checked with a standard |
| Speed | 3–9 min per sample | Depends on sample type and nitrogen content |
| Display / storage | 10.1-inch touchscreen, 16 GB, 5 million results | Historical results retrievable by date and sample ID |
| Power / weight / size | ≤1500 W, 16 kg, 405 × 465 × 690 mm | Bench instrument; needs a fume hood for reagent handling |
The repeatability figure deserves attention because it is the one a laboratory promises to clients. It describes agreement between parallel digestions of the same sample — which means it includes digestion variation, not only analyser variation. A laboratory that cannot digest reproducibly will not achieve 0.5% regardless of the analyser.
Reagents the laboratory must supply
The analyser handles the analysis, not the chemistry. Consumables the buyer must budget for include concentrated sulphuric acid, potassium sulphate and copper sulphate for digestion, sodium hydroxide for distillation, boric acid with indicator for absorption, and standard acid for titration. A digestion block is also separate equipment, and a laboratory without one should include it in the purchase rather than discover it at commissioning.
Beyond food: soil, feed and industrial samples
Kjeldahl nitrogen is not only a protein measurement. The same chemistry serves soil cation exchange capacity determination under LY/T 1243 and acrylonitrile content in rubber under SH/T 1157.2, and it is standard for feed and fertiliser nitrogen. A laboratory serving agricultural clients will find the 0.1–240 mg range covers most of those matrices, with the sample weight limit being the constraint for low-nitrogen soils.
Where an automatic analyser is the wrong purchase
- Very low sample numbers. Below about five samples a day, a semi-automatic instrument plus manual titration is cheaper and adequate.
- Samples needing a different nitrogen method. Dumas combustion is faster for high-throughput feed testing; Kjeldahl remains the reference method and is preferred where the standard specifies it.
- Laboratories without digestion capacity. The analyser cannot compensate for a digestor that cannot reach or hold temperature.
How HENG MEI supports Kjeldahl work
HENG MEI Intelligent Manufacturing supplies the HM-DN2 with full automatic reagent addition, distillation, absorption, static titration, emptying and rinsing, and 10.1-inch control, alongside the dynamic-titration HM-DN3 and the semi-automatic HM-DN1 for lower workloads. All three share the same digestion chemistry, so a laboratory can stage its investment.
To match a model to a workload, tell our sales contact the samples per day, the matrices, whether duplicates are routine, and whether the reporting standard names Kjeldahl or permits Dumas. Those four facts decide between semi-automatic, static and dynamic configurations.
Contact: Kyra, Sales Contact, Kyra@hmdzkj.cn, +86 13792642016 (Tel / WhatsApp / WeChat).
Address of this article:https://www.kjhm.net/news/automatic-kjeldahl-analyser-static-dynamic-titration.html
Related News
-
15 / 2026-06
Reliable Sterility Testing Pump Manufacturers & Brand Guide
Learn more
-
09 / 2026-11
Animal Disease Tester Brands Compared: IDEXX, Bio-Rad, HENG MEI
Learn more
-
16 / 2026-11
Fluorometric Nucleic Acid Quantification Guide | HENG MEI Intelligent Manufacturing
Learn more
-
13 / 2026-11
HENG MEI Intelligent Manufacturing Automated Portable Raman Analysis for Food Screening
Learn more
Current
Location: