Product Overview
The HM-P32 African swine fever detection equipment is mainly used to run real-time fluorescence quantitative PCR experiments. Through fluorescence excitation and collection, the experimental process is monitored in real time, and the experimental data is plotted into fluorescence curves, which are presented in real time on the instrument display interface. The experimental data is fitted and analyzed, and finally a PDF format experimental report can be generated.
Fast and accurate qualitative and quantitative analysis of target nucleic acids contained in biological samples taken from other animals (such as blood, body fluids, etc.), as well as standard curve, melting curve, genotyping analysis, etc. for special samples.
Basic Principles
The principle of HM-P32 African swine fever detection equipment: Polymerase chain reaction (PCR) is a molecular biology technology used to amplify specific DNA fragments. It can be seen as a special DNA amplification in vitro, consisting of several steps such as high-temperature denaturation, low-temperature annealing, and warm extension to form a temperature cycle, allowing the target gene to be rapidly amplified. It has strong specificity, high sensitivity, simple operation, and time-saving characteristics, and is a major innovation in gene amplification technology. PCR technology can specifically amplify extremely small amounts of target DNA by millions of times, greatly improving the analysis and detection capabilities of DNA molecules.
Real time fluorescence quantitative PCR technology is a method that adds corresponding fluorescent dyes or fluorescent labeled probes on the basis of conventional PCR. During the PCR reaction process, the entire PCR process is detected in real time through changes in fluorescence signals. The total amount of products after each polymerase chain reaction (PCR) cycle is monitored using fluorescent chemicals, and specific DNA sequences in the test sample are quantitatively analyzed using internal or external reference methods.
Basic structure
African swine fever detection equipment is a real-time detection instrument, mainly composed of gene amplification thermal cycle system, fluorescence real-time detection system, microcircuit control system, computer and application software. Two core functional modules are the thermal cycling system and the fluorescence real-time detection system. The working principle of the gene amplification thermal cycle system is basically the same as that of traditional gene amplification instruments, using semiconductor heating and cooling to complete the thermal cycle process. The fluorescence detection system mainly consists of fluorescence excitation components, light signal transmission components, fluorescence detection components, and control systems.
Application area
Basic scientific research
Pathogen detection
Animal epidemic detection
Meat product adulteration
□ Genetically modified testing
Food hygiene and quarantine
Drug development and rational use of drugs
Gene expression
Water monitoring
Technical Parameter
product name | YP-P16 | YP-P32 | YP-P48 |
appearance | |||
External dimensions | 235mm*385mm*175mm(Width * depth * height) | ||
weight | 5.6kg | 5.7kg | 5.8Kg |
Electrical parameters | ~220V/50Hz,255W | ||
data interface | USB 2.0 *2(Two on the right) | ||
Environmental parameters | |||
condition | temperature:10-30℃,humidity:20%~80% | ||
Transportation and storage conditions | temperature:-20~55℃,humidity:20%~80% | ||
altitude | <2500meter | ||
noise level | a-weighting,<60dB | ||
Sample parameters | |||
sample size | 16*0.2mL | 32*0.2mL | 48*0.2mL |
Test tube type | Single tube, eight row test tube | ||
Sample volume | 15-100uL | ||
Temperature | |||
Heating/cooling method | Semiconductor heating/cooling | ||
temperature range | 4℃-99℃ | ||
maximum rate | ≥3.5℃/s | ||
Average heating rate | ≥2.5℃/s | ||
Maximum cooling rate | ≥3.5℃/s | ||
Average cooling rate | ≥2.5℃/s | ||
precision | ≤±0.01℃ | ||
Temperature accuracy | ≤±0.1℃ | ||
Temperature uniformity | ≤±0.3℃ | ||
optical characteristics | |||
Number of fluorescence detection channels | 2 channels | 4 channels | 4 channels |
Light emitting devices | High-Brightness LED | ||
Daylighting device | High sensitivity, high signal-to-noise ratio photodiode | ||
Adapting probes or dyes | first channel:470/520 FAM,SYBR Green | ||
Second channel:530/570 HEX,JOE,VIC | |||
Third channel:580/610 ROX,CY3.5,Texas-Red | |||
Fourth channel:630/670 CY5 | |||
detection sensitivity | 1 copy | ||
Linear detection range | 100-1010 copies | ||
Linear correlation coefficient | ≥0.999 | ||
Channel crosstalk | No crosstalk | ||
Detect repeatability | ≤1.0% |