Product Introduction:
The full-function differential scanning calorimeter is a touch screen type that can perform glass transition temperature tests, phase transition tests, melting and thermal enthalpy tests, product stability, and oxidation induction period tests.
Comply with the following standards:
GB/T 19466.2–2004/ISO 11357-2:1999 Part 2: Determination of glass transition temperature
GB/T 19466.3–2004/ISO 11357-3:1999 Part 3: Determination of melting and crystallization temperature and enthalpy
GB/T 19466.6-2009/ISO 11357-3:1999 Part 6: Determination of oxidation induction time and oxidation induction temperature
Technical parameters:
1. Temperature range: -150~500℃ (minimum temperature determined according to refrigeration equipment)
2. Temperature resolution: 0.01℃
3. Temperature fluctuation: ±0.1℃
4. Temperature repeatability: ±0.1℃
5. Heating rate: 0.1~100℃/min
6. Constant temperature time: recommended <24h
7. Temperature control method: heating, constant temperature (automatic program control)
8.DSC range: 0~±600mW
9.DSC resolution: 0.01mW
10.DSC sensitivity: 0.01mW
11. Working power: AC220V/50Hz or customized
12. Optional cooling equipment:
- Compressed air cooling (-AR): RT ... 500℃
- Mechanical refrigeration (-MR): -40 ... 500℃
- Liquid nitrogen refrigeration (-LN): -150 ... 500℃
(Can be connected to multiple cooling devices individually or simultaneously)
13. Atmosphere control gas: nitrogen, oxygen (automatic switching of the instrument)
14. Gas flow: 0-300mL/min
15. Gas pressure: 0.2MPa
16. Display mode: 24-bit color, 8-inch LCD touch screen display
17. Data interface: standard USB interface
18. Parameter standard: equipped with standard materials (indium, tin, zinc), users can calibrate the temperature by themselves
Technical features:
1. Industrial-grade 8-inch touch screen, rich display information.
2. Brand-new furnace structure, can be matched with a variety of refrigeration devices.
3. USB communication interface, strong versatility, reliable and uninterrupted communication, support self-recovery connection function.
4. Automatically switch two-way atmosphere flow, fast switching speed, short stabilization time. At the same time, add one protective gas input.
5. The software is simple and easy to operate.