Product Introduction
The HM-T04L is a benchtop multi-parameter water quality analyzer manufactured by HM Instruments, designed to simultaneously measure four critical water quality indicators: Chemical Oxygen Demand (COD), ammonia nitrogen (NH₃-N), total phosphorus (TP), and total nitrogen (TN). Built on the rapid digestion spectrophotometry method specified in HJ/T 399-2007, this instrument delivers reliable and repeatable results for environmental monitoring laboratories.
The analyzer runs on an Android 7.1.1 smart operating system with a 10.1-inch high-definition color LCD touchscreen, providing an intuitive graphical interface for method selection, data analysis, and result visualization. Its precision colorimetric cell design ensures consistent light source alignment, eliminating errors caused by light source variation and delivering more accurate detection results. The integrated tube and dish colorimetric system supports both φ16mm digestion tubes and standard cuvettes (10mm, 30mm, 50mm), offering flexibility for different sample volumes and concentration ranges.
Bundled with a 6-hole rapid digestion instrument, the HM-T04L enables efficient sample preparation — heating up to 165°C within 10 minutes with intelligent PID temperature control. The complete system is suitable for industrial wastewater, municipal sewage, and surface water testing across industries including petrochemical, metallurgy, papermaking, brewing, and pharmaceuticals.
Applications
- Industrial Wastewater Monitoring: COD and nutrient analysis for petrochemical, metallurgical, and coking plant effluent
- Municipal Sewage Treatment: Routine monitoring of ammonia nitrogen, total phosphorus, and total nitrogen in treatment facilities
- Surface Water Assessment: Water quality evaluation for rivers, lakes, and reservoir basins
- Pharmaceutical and Brewing Industries: Process water and wastewater discharge compliance testing
- Papermaking and Chemical Plants: High-range COD measurement for production wastewater
- Environmental Regulatory Compliance: Data upload to environmental supervision platforms via LAN/WiFi
- Aquaculture Water Management: Nutrient level monitoring for fish farming and aquatic ecosystems
Features
- Android 7.1.1 intelligent operating system with fast processing speed and strong stability
- 10.1-inch high-definition color LCD touchscreen with bilingual Chinese/English interface
- Precision colorimetric cell design with consistent light source for more accurate results
- Imported ultra-high brightness LED light source with automatic intensity adjustment and calibration
- Dual colorimetric support: φ16mm tubes and 10mm/30mm/50mm cuvettes
- Wireless connectivity via WiFi, RJ45, and mobile hotspot for data transmission
- Multi-functional sample management with bilingual naming and 8 million data set storage
- Built-in cloud monitoring platform for direct data upload to environmental authorities
- Integrated thermal line printer with customizable print content including QR codes
- Optional 6Ah rechargeable lithium battery for portable outdoor field testing
- Built-in tutorial videos for guided step-by-step operation
- Firmware upgradeable for future method and feature updates
Main Parameters
| Parameter | Detection Method | Standard | Range | Detection Limit |
|---|---|---|---|---|
| COD | Rapid Digestion Spectrophotometry | HJ/T 399-2007 | 0–15,000 mg/L (segmented) | 3 mg/L |
| Ammonia Nitrogen | Nessler Reagent / Salicylic Acid Spectrophotometry | HJ 535-2009 / HJ 536-2009 | 0.2–200 mg/L / 0.025–25 mg/L | — |
| Total Phosphorus | Ammonium Molybdate Spectrophotometry | GB 11893-89 | 0–30 mg/L (segmented) | 0.02 mg/L |
| Total Nitrogen | Alkaline Potassium Persulfate Digestion UV Spectrophotometry | HJ 636-2012 | 0–200 mg/L (segmented) | 0.5 mg/L |
Technical Parameters
| Parameter | Value |
|---|---|
| Indication Error | ≤ ±5% |
| Instrument Stability | < 0.5% |
| Instrument Repeatability | < 0.5% |
| Photochemical Stability | ≤ 0.002A drift within 20 min (100,000-hour lifespan) |
| Colorimetric Method | Tube (φ16mm) + Cuvette (10mm, 30mm, 50mm) |
| Operating System | Android 7.1.1 |
| Display | 10.1-inch HD color LCD touchscreen |
| Interface Language | Chinese / English |
| Network Interface | USB 2.0, HDMI, WiFi, Bluetooth, Hotspot, RJ45 |
| Cloud Platform | Built-in regulatory monitoring platform |
| Printer | Thermal line printer |
| Data Storage | 8 million sets |
| Data Export | Excel format via USB |
| Instrument Size | 444 × 303 × 173 mm |
| Instrument Weight | 5.2 kg |
| Rated Power | 10W |
| Power Supply | AC 220V ±10% / 50Hz |
| Environmental Temperature | 5–40°C |
| Environmental Humidity | < 85% RH (no condensation) |